28 Parameters

915 KF Ti-Touch

■■■■■■■■

307

Filling rate

Rate at which the dosing cylinder is filled.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

maximum
The maximum filling rate depends on the cylinder volume (see Chapter
31, page 338)
.

28.6.3.2

Control device

The control device the determination is carried out with is selected under
[Control device]. Control devices are defined under System

Device

manager.

NOTE

This button will only be shown when, in addition to the Ti-Touch, a
Sample Processor has been configured as control device.

Control device

Selection of the control device from the list of devices. Only those devices
are displayed which are able to carry out the command.

Selection

Selection of configured control devices

28.6.3.3

Dosing device

The parameters for the dosing device are edited under [Dosing device].

Dosing device

Selection of the MSB connector the dosing device is connected to. The
selection does not depend on the number of MSB connectors of the con-
trol device.

Selection

1 | 2 | 3 | 4

Default value

1

Titrant

Selection of the titrant from the titrant list. We recommend always select-
ing the titrant. This ensures that the correct data (titer, etc.) is always used
for the calculation. Titrants are defined under System

Titrants. You

can also enter a name which is not contained in the titrant list. When a
determination is started there is a check whether the solution is contained
in the list.