NOTE

A line set active is not being reset automatically, not even at the end of
the determination.

Entering the bit pattern:

0 = line inactive

1 = line active

* = retain line status

p = set pulse

The output lines are always numbered from right to left, i.e. with the sig-
nal *************1 line 0 is set. With a pulse, the length is set to 200 ms.
If you wish to set pulses with other lengths, you have to define a corre-
sponding template.

Entry

Bit pattern containing exactly 14 characters
or a max. of 24 characters for the name of
the template

Default value

**************

Selection

Selection of the templates defined

16.5.3

Sample data

Main dialog: Edit parameters

Method options Sample data

You can define the following method-specific settings in the dialog
Method options / Sample data:

Modify the designation for the sample identifications

Define a fixed sample size

Define limits for the sample size
If you define limits, these will be monitored at the start and at the end
of the determination.

Below you will find an example demonstrating the effects on the main
dialog of the settings in this dialog (see "Modifying sample data", page
165)
.

Identification 1, Identification 2

on | off (Default value: on)

If this option is activated, then the input field for the sample identification
in the main dialog will be displayed.

Designation

A designation of your own choosing can be defined for the input field for
each method.