2 measuring inputs, 1 polarizer, 2 temperature – Metrohm 915 KF Ti-Touch User Manual

Page 370: 2 measuring inputs 32.2.1, Polarizer, Temperature

Advertising
background image

32.2 Measuring inputs

■■■■■■■■■■■■■■■■■■■■■■

356

■■■■■■■■

915 KF Ti-Touch

32.2

Measuring inputs

The measuring cycle is 100 ms for all measuring modes.

32.2.1

Polarizer

A measuring input (Pol.) for polarizable electrodes.

Measuring mode
Ipol

Determination with adjustable polarization current.

Polarization
current

–120 - +120 µA (increment: 1 µA)
–125 - –121 µA / +121 - +125 µA: non-guaranteed values, dependent
on reference voltage +2.5 V

Measuring
range

–1,200 - +1,200 mV

Resolution

0.1 mV

Measuring
accuracy

±0.2 mV
(±1 digit, without sensor error, under reference conditions)

Measuring mode
Upol

Determination with adjustable polarization voltage.

Polarization
voltage

–1,200 - +1,200 mV (increment: 10 mV)
–1,250 - –1,210 mV / +1,210 - +1,250 mV: non-guaranteed values,
dependent on reference voltage +2.5 V

Measuring
range

–120 - +120 µA

Resolution

0.01 µA

Measuring
accuracy

32.2.2

Temperature

A measuring input (Temp.) for temperature sensors of the Pt1000 or NTC
type with automatic temperature compensation.

R (25 °C) and B value can be configured for NTC sensors.

Measuring range

Pt1000

–150 - +250 °C

NTC

–5 - +250 °C
(For an NTC sensor with R (25 °C) = 30,000

Ω and B (25/50) = 4,100 K)

Resolution

Pt1000

0.1 °C

NTC

0.1 °C