21 Determination run

915 KF Ti-Touch

■■■■■■■■

189

21 Determination run

This chapter provides a description of how to carry out a determination,
i.e. how to process a single sample or a sample series. The current sample
data is used for calculations.

21.1

Carrying out a single determination

When you carry out a determination, you can enter the sample data in
various ways (see Chapter 19, page 178). The following instructions
describe how to enter the sample data in the main dialog.

To carry out a single determination, proceed as follows:

1 Load the method

See Chapter "Loading a method", Page 156.

2 Prepare the sample

NOTE

Calculate the amount of the sample in such a way that a titrant
consumption of 10…90 % of the cylinder volume will result.

Weigh in or measure the sample in a sample vessel.

Add solvent.

Place the sample vessel on the stirrer.

Immerse the electrode and buret tip in the solution.

3 Enter the sample size

In the main dialog, tap on the input field Sample size.
The number editor opens.

Enter the sample size and confirm with [OK].

4 Start the determination

Tap on the fixed key [

].