9.2 Reagent monitoring

■■■■■■■■■■■■■■■■■■■■■■

68

■■■■■■■■

915 KF Ti-Touch

at the end of the determination.

Number of determ.

The number of determinations to be carried out with a certain amount of
reagent depends on the type of sample and its amount.

Input range

1 - 999

Selection

off

Default value

off

Working life

Working life of the reagent.

Input range

1 - 999 days

Selection

off

Default value

off

Volume

Volume of titrant dosed.

Input range

1.0 - 999.9 mL

Selection

off

Default value

off

[Status]

Display the status overview of the current values of the reagent monitor-
ing.

[Reagent replacement]

Edit the parameters for the reagent exchange.

Dialog "Reagent monitoring / Status"
The current reagent monitoring values are displayed in this dialog.

[Reset]

Reset the values to zero.

Dialog "Reagent monitoring / Reagent replacement"
The parameters for the reagent exchange are defined in this dialog.

Reagent replacement

The reagent can either be exchanged manually or automatically.

Selection

manual | auto

Default value

manual

manual
If a monitored parameter has reached the limit set, a message is being
displayed. Then the reagent has to be exchanged manually.