27 Manual control

915 KF Ti-Touch

■■■■■■■■

243

Continuous dosing is paused with [Hold]. The label changes to
[Continue]. This can be used to continue the run.

3 Fill the cylinder

With the default settings (see "Fill automatic.", page 244) the dos-
ing cylinder is filled automatically.

Otherwise:

Tap on [Fill].

The dosing cylinder is filled. The displayed volume value will be reset
to 0.0000 mL.

Parameter description

Volume

Volume which is dosed.

Input range

0.00000 - 99999.9 mL

Default value

0.10000 mL

Dosing rate

Rate at which it is dosed.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 31, page 338)
.

Filling rate

Rate at which the dosing cylinder is filled.