2 calculations (calc live), Calculations (calc live) – Metrohm 915 KF Ti-Touch User Manual

Page 305

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

28 Parameters

915 KF Ti-Touch

■■■■■■■■

291

Precision

Setting, with which accuracy the result is used in additional calculations.

Selection

Round | Truncate | Full precision

Default value

Round

Round
The result is rounded to the defined number of decimal places (com-
mercial rounding, in accordance with the US Pharmacopeia USP). If the
digit at the first dropped decimal place is 1, 2, 3 or 4, then it will be
rounded off; if this digit is 5, 6, 7, 8 or 9, then it will be rounded up.
Negative digits will be rounded in accordance with their amount, i.e.
away from zero.
Truncate
The result is cut to the number of decimal places defined.
Full precision
The result is used with full accuracy (floating point number either in
"single precision" (32 bit) or in "double precision" (64 bit), according to
the standard IEEE 754).

28.4.2

Calculations (CALC LIVE)

The command CALC LIVE can be used to define a calculation, the result
of which will be shown in the live display during a titration or measure-
ment. The current volume or measured value, respectively, is used for this
purpose. This is helpful, e.g. for Karl Fischer titrations, in order to be able
to track the water content directly during the titration.

CAUTION

The command must be inserted directly before the corresponding titra-
tion or measuring command.

This command is identical with the CALC command except for the follow-
ing differences:

The number of variables is limited.

There are no result options.

The result cannot be monitored.

The result variable cannot be modified.

The result will only be displayed in the live display. It appears neither in
the results dialog nor in a report.

Result name

With live calculations no custom result name can be defined, the designa-
tion LR cannot be modified.

Advertising