Metrohm 915 KF Ti-Touch User Manual

Page 160

Advertising
background image

15.2 Custom result templates

■■■■■■■■■■■■■■■■■■■■■■

146

■■■■■■■■

915 KF Ti-Touch

Display result

on | off (Default value: on)

If you deactivate this parameter, the result is neither displayed in the result
dialog nor printed in the result report. This can be advisable for intermedi-
ate results.

Save result in result table

on | off (Default value: off)

The calculated result can be saved in the result table. This may be advisa-
ble if e.g. the results of all determinations carried out on a particular day
are to be displayed clearly. A maximum of nine results from a determina-
tion can be saved in the result table.

Precision

Setting, with which accuracy the result is used in additional calculations.

Selection

Round | Truncate | Full precision

Default value

Round

Round
The result is rounded to the defined number of decimal places (com-
mercial rounding, in accordance with the US Pharmacopeia USP). If the
digit at the first dropped decimal place is 1, 2, 3 or 4, then it will be
rounded off; if this digit is 5, 6, 7, 8 or 9, then it will be rounded up.
Negative digits will be rounded in accordance with their amount, i.e.
away from zero.
Truncate
The result is cut to the number of decimal places defined.
Full precision
The result is used with full accuracy (floating point number either in
"single precision" (32 bit) or in "double precision" (64 bit), according to
the standard IEEE 754).