21.2 Performing a sample series

■■■■■■■■■■■■■■■■■■■■■■

190

■■■■■■■■

915 KF Ti-Touch

Sequence of the determination

After a determination is started, a check is made as to

– whether all of the devices required are connected and ready for

operation.

– whether all titrants and sensors are present.
– whether the monitoring intervals for titrants, sensors and com-

mon variables are still valid.

if a sample assignment table is activated (see Chapter 15.1.2,

page 141):
whether the entered sample identification is defined in the sam-
ple assignment table. If this is the case, then the assigned
method will be loaded automatically and the actual determina-
tion will be started.

21.2

Performing a sample series

You can use the following functions with sample series:

Statistics for multiple determinations
The statistics calculations are enabled in the method options (see
Chapter 16.5, page 159)
.

Sample table
You can enter the sample data for an entire sample series in the sam-
ple table (see Chapter 20, page 181).

Sequence of the sample series

After a determination is started, a check is made as to

– whether all of the devices required are connected and ready for

operation.

– whether all titrants and sensors are present.
– whether the monitoring intervals for titrants, sensors and com-

mon variables are still valid.

if a sample assignment table is activated (see Chapter 15.1.2,

page 141):
whether the entered sample identification appears in the sample
assignment table. If this is the case, then the assigned method
will be loaded automatically and the actual determination will be
started. If a different method has been entered in the sample
table, then it will be ignored.

At the end of the determination,

– the line in the sample table is deleted.
– the next determination is started (if you are working with the

autostart function enabled (see Chapter 17, page 170)).