Metrohm 915 KF Ti-Touch User Manual

Page 68

Advertising
background image

8.4 Dosing unit

■■■■■■■■■■■■■■■■■■■■■■

54

■■■■■■■■

915 KF Ti-Touch

NOTE

These parameters are important for the correct execution of the Pre-
paring
(PREP command) and Emptying (EMPTY command) functions,
because the volumes of the tubing connections are taken into account.

Dosing port 1

Port

Port to be used as dosing port 1 for the PREP and EMPTY (see Figure
10, page 56)
functions.

Selection

Port 1 | Port 2 | Port 3 | Port 4

Default value

Port 1

Length

Length of the tubing.

Input range

0.0 - 999.9 cm

Default value

40.0 cm
The setting 0.0 means that this tubing will neither
be rinsed nor emptied.

Diameter

Diameter of the tubing.

Input range

0.0 - 9.9 mm

Default value

2.0 mm

Dosing port 2

Port

Port to be used as dosing port 2 for the PREP and EMPTY functions (see
Figure 10, page 56)
.

Selection

Port 1 | Port 2 | Port 3 | Port 4

Default value

Port 3

Length

Length of the tubing.

Input range

0.0 - 999.9 cm

Default value

0.0 cm
The setting 0.0 means that this tubing will neither
be rinsed nor emptied.