11 Device manager

915 KF Ti-Touch

■■■■■■■■

99

A maximum of 16 rack positions can be defined as special beakers for
each sample rack. The dialog Edit rack data / Special beakers shows a
list of all special beakers with their assigned rack positions.

[Edit]

Edit the data of the selected special beaker, see the following.

You can edit the data of the selected special beaker in the dialog Special
beaker / Edit
.

Rack position

Number of rack position for selected special beaker. Any rack position can
be defined as a special beaker. It is, however, preferable to set them at
high rack positions in order to be able to begin sample series at rack posi-
tion 1. Rack positions defined as special beakers can no longer be used as
sample positions.

Input range

0 - 'maximum number of rack positions'

Work pos. tower 1/2

Work position for the selected special beaker. One specific work position
can be defined for tower 1 and tower 2.

Input range

0 - 'maximum stroke path' mm
A lift position of 0 mm corresponds to the "home
position", i.e. the lift is located at the upper stop
position. The maximum stroke path is defined in the
properties of the tower (see "Maximum stroke
path", page 88)
.

Beaker radius

Actual radius of the selected special beaker.

This beaker radius may not be less than the minimum beaker radius
defined in the tower properties (see "Min. beaker radius," page 88). If the

Advertising