Metrohm 915 KF Ti-Touch User Manual

Page 110

Advertising
background image

11.6 Sample racks

■■■■■■■■■■■■■■■■■■■■■■

96

■■■■■■■■

915 KF Ti-Touch

You can edit the data of the selected rack in the dialog Sample rack /
Edit rack data
.

Beaker radius samples

Actual radius of the sample vessels at the general sample positions of the
rack.

This beaker radius may not be less than the minimum beaker radius
defined in the tower properties (see "Min. beaker radius," page 88). If the
lift is to be moved to the work position, then these two values will be
compared with one another.

Input range

1.0 - 100.0 mm

Selection

off

Default value

off

off
No check takes place.

Beaker sensor

Each time a sample position is to be moved to with the MOVE command,
the beaker sensor checks whether a vessel is present. In the MOVE com-
mand, you define the action that takes place if the beaker sensor does not
detect a vessel at the position being moved to.

Selection

Tower | Robotic arm | off

Default value

off

Robotic arm
A Swing Head with beaker sensor must be mounted. In addition, a
suitable work position must be defined for the lift, so that the robotic
arm touches the sample vessel. The work position is moved to for the
purpose of beaker detection.
off
No check takes place.

Rack offset

The rack offset is a production-related tolerance value between the upper
and lower sections of the rack. This value is determined when the rack is
adjusted and is displayed here. It can be changed if necessary.

Input range

–10.00 - 10.00 °

Default value

0.00 °

[Adjust rack]

Adjust the rack (see Chapter 11.6.2, page 100).

[Lift pos. tower 1]/[Lift pos. tower 2]

Define rack specific lift positions (see "Lift positions Tower 1/2", page
97)
.