28.3 Evaluations (EVAL)

■■■■■■■■■■■■■■■■■■■■■■

278

■■■■■■■■

915 KF Ti-Touch

Fixed quantity

Selection of the quantity to which the associated value is interpolated
from the measuring point list.

Selection

Measured value | Time | Volume

Default value

Measured value

Fixed EP1 at

The fixed endpoint must lie between the first and the final entry in the
measuring point list.

The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the 32 Technical specifications chapter, page 355.

Measured value (measuring mode Ipol):
Input range

–2,000.0 - 2,000.0 mV

Selection

off

Default value

off

Measured value (measuring mode Upol):
Input range

–200.00 - 200.00 µA

Selection

off

Default value

off

Measured value (measuring mode T):
Input range

–20.0 - 150.0 °C

Selection

off

Default value

off

Time:
Input range

0.0 - 999,999.9 s

Selection

off

Default value

off

Volume:
Input range

0.00000 - 9,999.99 mL

Selection

off

Default value

off

Fixed EP2 at…Fixed EP9 at

See Fixed EP1 at.