28.1 Volumetric Karl Fischer titrations (KFT)

■■■■■■■■■■■■■■■■■■■■■■

268

■■■■■■■■

915 KF Ti-Touch

NOTE

This parameter must be activated if commands follow which must not
run simultaneously with the conditioning.

Example: Emptying the titration cell followed by adding new working
medium.

28.1.6

Cell

The monitoring of the reagent used is activated or deactivated under
[Cell].

Reagent monitoring

Selection of the reagent from the list of reagents. The selection depends
on the titration mode. Reagents are defined under System

Reagents.

Selection

Selection of configured reagents | off

Default value

off

off
The reagent monitoring is not active.

28.1.7

Control device

The control device the determination is carried out with is selected under
[Control device]. Control devices are defined under System

Device

manager.

NOTE

This button will only be shown when, in addition to the Ti-Touch, a
Sample Processor has been configured as control device.

Control device

Selection of the control device from the list of devices. Only those devices
are displayed which are able to carry out the command.

Selection

Selection of configured control devices