27 Manual control

915 KF Ti-Touch

■■■■■■■■

241

NOTE

If you exit the dialog with the fixed key [

], the dosing cylinder

will be filled automatically.

Parameter description

Dosing rate

Rate at which it is dosed.

Input range

0.01 - 166.00 mL/min

Selection

maximum | dynamic

Default value

maximum

maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 31, page 338)
.
dynamic
This setting can only be selected when the dosing cylinder is filled. The
dosing is being carried out faster and faster until the maximum dosing
rate is reached (starting with 1 mL/min and then doubling the dosing
rate every 1.5 s).

Filling rate

Rate at which the dosing cylinder is filled.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

maximum
The maximum filling rate depends on the cylinder volume (see Chapter
31, page 338)
.

27.3.2

Dosing fixed volumes

Manual control: Dosing

Dosing fixed vol.