3 preparing, Preparing, Figure 12 – Metrohm 915 KF Ti-Touch User Manual

Page 258: Live display "preparing the dosing unit

Advertising
background image

27.3 Dosing

■■■■■■■■■■■■■■■■■■■■■■

244

■■■■■■■■

915 KF Ti-Touch

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

maximum
The maximum filling rate depends on the cylinder volume (see Chapter
31, page 338)
.

Fill automatic.

on | off (Default value: on)

If this parameter is activated, then the dosing cylinder is being automati-
cally filled at the end of the dosing.

27.3.3

Preparing

Manual control: Dosing

Prepare

The Prepare function is used to rinse and fill the cylinder and tubings of
the exchange or dosing unit air bubble-free. You should carry out this
function before the first determination or once per day.

If the titrant is selected (see "Selecting the dosing function", page 238),
the parameters defined for the titrant for preparing/emptying and the tub-
ing parameters will be used (dosing unit only). If the titrant is not selected,
default parameters will be used (see Chapter 31.1.2, page 338 and Chap-
ter 31.2.2, page 339)
.

The following two figures show the live display of an exchange unit and a
dosing unit:

Figure 12

Live display "Preparing the dosing unit"

Advertising