15.1 Sample data

■■■■■■■■■■■■■■■■■■■■■■

142

■■■■■■■■

915 KF Ti-Touch

In the sample assignment table, a particular method is assigned to a sam-
ple identification. In this way, you ensure that your samples will be pro-
cessed with the correct method; mix-ups are not longer possible. When
you start a determination, you need only enter the sample identification;
the method is loaded automatically.

[New]

Add a new sample assignment to the list, see following chapter.

[Delete]

Delete the selected sample assignment.

[Edit]

Edit the selected sample assignment, see following chapter.

Editing the sample assignment

Identification

Identification of the sample.

NOTE

You can place an * as a wildcard at the beginning or end of the charac-
ter string. Doing so allows you to prefix or suffix a sequential number,
for example, to the identification, which will be ignored when methods
are being assigned.

Entry

24 characters maximum

Selection

Selection of defined identifications in the
sample identification list

Memory

Memory location the method is loaded from. All memory locations are
selectable, even if they are currently not accessible.

Selection

Internal memory | External memory 1 | Exter-
nal memory 2
| Shared memory

Default value

Internal memory

Shared memory
Shared directory in the network.

Method

Method that is loaded when a sample with the above-specified identifica-
tion is processed. You can also enter a method that does not yet exist in
the selected memory. When a determination is started there is a check
whether the method is available.