Metrohm 915 KF Ti-Touch User Manual

Page 264

Advertising
background image

27.6 USB Sample Processor

■■■■■■■■■■■■■■■■■■■■■■

250

■■■■■■■■

915 KF Ti-Touch

NOTE

Note that the rack has to be set to a valid position for all lift move-
ments! Otherwise a corresponding error message will be displayed.

Move to a valid rack position with [Previous] or [Next]. Alternatively,
you can carry out a MOVE command.

[Home position]

Move the lift to the home position (corresponds to lift position 0 mm).

[Work position]

Move the lift to the work position of the current rack position or the
external position.

The work position can directly be defined here (see "Assigning lift posi-
tions", page 252)
or in the device manager
(see "Lift positions Tower
1/2", page 97)
.

[

]

Continuously move the lift upwards as long as the button is pressed
down. The lift rate used is defined in the dialog Sample Processor / Lift.

[

]

Continuously move the lift downwards as long as the button is pressed
down. The lift rate used is defined in the dialog Sample Processor / Lift.

[Previous]

Move to the previous rack position.

[Next]

Move to the next rack position.

[Reset rack]

Initialize the rack. This function is identical with the command RACK. The
following actions are carried out:

The lift is moved upwards.

The sample rack is rotated to the starting position.

The rack code of the rack attached is being read out and the corre-
sponding rack data is transferred to the Sample Processor.

The sample variable is being reset to the value 1.

[Pump]

Switch on and off the pumps mounted or connected to the tower.

[Lift]

Move the lift to any or a specific lift position and assigning lift positions
(see Chapter 27.6.1, page 251).

Advertising