Metrohm 915 KF Ti-Touch User Manual

Page 227

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

23 Results and more determination data

915 KF Ti-Touch

■■■■■■■■

213

EVAL commands:
All evaluations are carried out again. You can also insert new evalua-
tions retroactively.

CALC command:
All calculations are carried out again. You can also define new calcula-
tions retroactively.
If you subsequently modify the titrant in a method, then the variable
TITER of the "new" titrant will be used. If a result is assigned to the var-
iable TITER in a calculation, then the titer of the "new" titrant will also
be overwritten.

If a determination has been recalculated, this will be documented in the
determination properties (see Chapter 23.4, page 203).

Recalculation/reevaluation of loaded determinations
With loaded determinations, it is assumed that they were carried out a
long time ago and that the titrant data and common variables of the
system no longer correspond to those of the determination. This is why
the most important titrant data of the titrants used and the list of com-
mon variables are stored with each determination. This data will be used if
you recalculate a loaded determination. If you wish to recalculate the
result with a corrected common variable or with a corrected titer, you
must therefore modify the respective values in the More determination
data / Local common variables
or More determination data / View
data
dialog.

If a result is assigned to the variable TITER or to a common variable, then
the titers or common variables which are currently valid in the system will
not be overwritten until after a query has been confirmed.

If you modify the titrant retroactively in the method, this will not be taken
into account.

The recalculated results of loaded determinations are not saved in the
result table.