2 editing a calculation – Metrohm 915 KF Ti-Touch User Manual
Page 301
■■■■■■■■■■■■■■■■■■■■■■
28 Parameters
915 KF Ti-Touch
■■■■■■■■
287
The calculation can now be edited further, e.g. define result limits,
see following chapter.
28.4.1.2
Editing a calculation
Result name
The result name is the text which will be shown in the result display and in
the report. The default name corresponds to the result variable.
Entry
max. 24 characters
Default value
R1…R9
Calc. formula R1...R9
Shows the calculation formula. A special editor is opened for the defini-
tion (see Chapter 28.4.3, page 293). The designation R1...R9 corre-
sponds to the result variable. Under [Result variable] this can later be
modified.
Entry
max. 100 characters
Default value
empty
Decimal places
Number of decimal places used to display the result.
Input range
0 - 5
Default value
2
Result unit
The result unit is displayed and saved together with the result.
Entry
max. 10 characters
Selection
% | mol/L | mmol/L | g/L | mg/L | mg/mL | mg/
100 g | ppm | g | mg | µg | mL | µL | mg/piece |
°C | µg/min | mL/min | µL/min
Default value
%