3 dosing a specified volume (add), 1 dosing parameters, Dosing a specified volume (add) – Metrohm 915 KF Ti-Touch User Manual

Page 320

Advertising
background image

28.6 Dosing and Liquid Handling

■■■■■■■■■■■■■■■■■■■■■■

306

■■■■■■■■

915 KF Ti-Touch

which is not contained in the titrant list. When a determination is started
there is a check whether the solution is contained in the list.

For buret units with integrated data chip, a check is made in the method
run to verify whether the correct titrant has been attached and whether
the type of dosing drive matches. For buret units without integrated data
chip, the cylinder volume and the type of dosing drive are checked. For
the selected titrant, the validity of the titer, the working life of the titrant
and the time interval for the GLP test of the buret unit are checked at the
start of the determination.

Selection

Selection of configured titrants | not defined

Default value

not defined

not defined
No check takes place. The command is carried out with default param-
eters (see Chapter 31.1.2, page 338 and Chapter 31.2.2, page 339).

Air inlet

This parameter can only be edited with Titrant = not defined.

Selection of the port via which air is aspirated. This setting is only relevant
for dosing units, in case of preparing an exchange unit, this setting is
ignored.

Selection

Port 1 | Port 2 | Port 3 | Port 4

Default value

Port 4

28.6.3

Dosing a specified volume (ADD)

You can dose a specified volume with the command ADD.

28.6.3.1

Dosing parameters

The parameters for the dosing are defined under [Dosing parameters].

Volume

Volume which is dosed.

Input range

0.00000 - 99999.9 mL

Default value

10.0000 mL

Dosing rate

Rate at which it is dosed.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 31, page 338)
.

Advertising