Metrohm 915 KF Ti-Touch User Manual

Page 144

Advertising
background image

13.4 System validation

■■■■■■■■■■■■■■■■■■■■■■

130

■■■■■■■■

915 KF Ti-Touch

Selection

Display message | Document message | Cancel
determination

Default value

Display message
For all three options it is documented in the deter-
mination data (see dialog More determination
data / Messages
), that the time interval has been
expired.

Display message
A message is displayed. You can select whether you want to continue
with the determination or cancel the run.
Document message
In the determination data it will be documented, that the time interval
has been expired.
Cancel determination
The determination is stopped.

Dialog "System validation / Test statistics"
In the dialog System validation / Test statistics, you can document
the statistical data for the last system validation.

Number (n)

Number of determinations carried out.

Input range

1 - 9999999999

Default value

empty

Mean value

Mean value out of single results.

Input range

–999999999.00000 - 999999999.00000

Default value

empty

s abs

Absolute standard deviation of the results.

Input range

0.00000 - 999999999.00000

Default value

empty

s rel

Relative standard deviation of the results.

Input range

0.00000 - 100.00000 %

Default value

empty %

d rel

Systematic deviation of the results.