15.3 Input lines

■■■■■■■■■■■■■■■■■■■■■■

148

■■■■■■■■

915 KF Ti-Touch

Signal name

Input signal

Function

Sample ready

***1****

waits for a stepping pulse, of, for instance, a con-
nected Sample Processor, as soon as it is ready.

[New]

Add a new input signal to the list, see following chapter.

[Delete]

Delete the selected input signal from the list.

[Edit]

Edit the data of the selected input signal, see following chapter.

15.3.1

Editing the input signal

List of input signals: Input signal

New / Edit

Signal name

Designation of the template.

Entry

24 characters maximum

Input signal

Entering the desired bit pattern:

Entering the bit pattern:

0 = line inactive

1 = line active

* = retain line status

The input lines are always numbered from right to left, i.e. with the signal
*******1 line 0 is expected to be active.

NOTE

We recommend masking lines that are of no interest or for which no
defined condition can be predicted with an asterisk (*).

Entry

Bit pattern of exactly 8 characters

Default value

********