27 Manual control

915 KF Ti-Touch

■■■■■■■■

247

The MSB connector and type of stirrer is indicated for each connected stir-
rer.

You can control a connected stirrer manually with the function [Stir].

Proceed as follows:

1 Set the stirring rate

Tap on the button [–] or [+].
Each time one of the buttons is tapped, the stirring rate is
reduced or increased by one level.

2 Switch on the stirrer

Tap on [Start].

The stirrer is started.

3 Switch off the stirrer

Tap on [Stop].

The stirrer is stopped.

Parameter description

Stirring rate

Setting the stirring rate. It can be set in steps of –15 to +15. The default
setting 8 corresponds to 1000 rpm. The formula for calculating the rota-
tional speed is specified in chapter 31.3, page 339.

The algebraic sign of the stirring rate changes the direction in which the
stirring is done. When the stirrer is viewed from above, this means:

"+": counterclockwise rotation

"–": clockwise rotation

Input range

–15 - 15

Default value

8

NOTE

If a stirrer is used in the determination run, it can nevertheless be con-
trolled manually at the same time. For example, you can use the man-
ual control to reduce the stirring rate of a stirrer which has been started
in the determination run.