3 monitoring the working life – Metrohm 915 KF Ti-Touch User Manual

Page 86

Advertising
background image

10.3 Monitoring the working life

■■■■■■■■■■■■■■■■■■■■■■

72

■■■■■■■■

915 KF Ti-Touch

Entry

8 digits maximum

Comment

Entry

24 characters maximum

[Working life]

Define the working life of the sensor (see Chapter 10.3, page 72).

10.3

Monitoring the working life

Sensor: Edit

Working life

In the dialog Edit sensor / Working life, you can define the time inter-
val after which the sensor must be replaced. If you do not wish to monitor
the working life, then you can enter only the date of manufacture for doc-
umentation purposes.

Start-up

Date on which the sensor was used for the first time.

Format: YYYY:MM:DD

Monitoring

on | off (Default value: off)

If this parameter is activated, then the working life will be monitored.

Working life

If you define a time interval for the working life, then the Expiry date will
be tracked automatically.

Input range

1 - 999 days

Default value

999 days

Expiry date

If you define an expiry date, then the Working life will be tracked auto-
matically.

Format: YYYY:MM:DD

Action

Selection of the action which is carried out when the time interval has
expired.

Advertising