28.1 Volumetric Karl Fischer titrations (KFT)

■■■■■■■■■■■■■■■■■■■■■■

270

■■■■■■■■

915 KF Ti-Touch

Electrode check

on | off (Default value: off)

For the following electrodes, an electrode check can be carried out:

pH electrodes

Metal electrodes

Ion-selective electrodes

A check is made that the electrode is properly connected and that no
short-circuit is present. The electrode check is carried out when this com-
mand is started.

Temp. measurement

Type of temperature measurement.

Selection

continuous | automatic | off

Default value

automatic

continuous
A temperature sensor must be connected. The temperature is mea-
sured continuously.
automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise, the temperature entered manually
will be used (see dialog of the titration and measuring parameters).
off
The temperature will not be measured. The temperature entered man-
ually is used (see dialog of the titration and measuring parameters).

28.1.9

Dosing device

The parameters for the dosing device are edited under [Dosing device].

Dosing device

Selection of the MSB connector the dosing device is connected to. The
selection does not depend on the number of MSB connectors of the con-
trol device.

Selection

1 | 2 | 3 | 4

Default value

1

Titrant

Selection of the titrant from the titrant list. We recommend always select-
ing the titrant. This ensures that the correct data (titer, etc.) is always used
for the calculation. Titrants are defined under System

Titrants. You

can also enter a name which is not contained in the titrant list. When a
determination is started there is a check whether the solution is contained
in the list.