28 Parameters

915 KF Ti-Touch

■■■■■■■■

277

Switch off automatically

on | off (Default value: on)

If this parameter is activated, the stirrer will be switched off automatically
at the end of the titration, measurement, etc.

28.3

Evaluations (EVAL)

Various additional evaluations (EVAL commands) can be carried out for
titrations and measurements. The evaluation always refers to the last titra-
tion or measurement prior to the EVAL command. Only those evaluation
commands which are available for the last titration or measurement prior
to the EVAL command can be inserted in the list of commands. If a titra-
tion or measuring command is deleted before the EVAL command, then it
will be shown in red in the list of commands, as the reference is missing.

EVAL commands can also be inserted in the method run at a later time
and the determination can be reevaluated ([Recalculate] function in the
results dialog).

Additional evaluations for KFT titrations
The following additional evaluations are possible:

EVAL FIX-EP (fixed endpoints)
Measured quantities:

– Measured value
– Time
– Volume

EVAL MIN/MAX (minimum/maximum)

EVAL RATE (average dosing rate)

Additional evaluations for measurements
The following additional evaluations are possible:

EVAL FIX-EP (fixed endpoints)
Measured quantities:

– Measured value
– Time

EVAL MIN/MAX (minimum/maximum)

EVAL BREAK (break points)

28.3.1

Fixed endpoint evaluation (EVAL FIX-EP)

For the fixed endpoint evaluation, the associated values are interpolated
from the measuring point list for one quantity (measured value, volume,
etc.). Up to nine fixed endpoints can be evaluated with one command.