28.7.2

Setting remote lines (CTRL)

The command CTRL can be used to define output signals at the remote
interface.

Control device

This parameter will only be shown when, in addition to the Ti-Touch, a
Sample Processor has been configured as control device. Control devices
are defined under System

Device manager.

Selection of the control device from the list of devices.

Selection

Selection of configured control devices

Remote Box

Selection of the MSB connector the Remote Box is connected to. The
selection does not depend on the number of MSB connectors of the con-
trol device. The remote signal defined is sent via this Remote Box.

Selection

1 | 2 | 3 | 4

Default value

1

Output signal

Selection of the signal out of the templates or entering the required bit
pattern. Templates are defined under System

Templates Output

lines.

NOTE

A line set active is not being reset automatically, not even at the end of
the determination.

Entering a bit pattern:

0 = line inactive

1 = line active

* = retain line status

p = set pulse

The output lines are always numbered from right to left, i.e. with the sig-
nal *************1 line 0 is set. With a pulse, the length is set to 200 ms.