8.3 Monitoring the working life

■■■■■■■■■■■■■■■■■■■■■■

50

■■■■■■■■

915 KF Ti-Touch

8.3

Monitoring the working life

Titrant: Edit

Working life

In the dialog Edit titrant / Working life, you can define the time inter-
val after which the titrant must be replaced. This is particularly important if
your titrant has a limited working life. If you do not wish to monitor the
working life, then you can enter only the date of manufacture for docu-
mentation purposes.

Preparation date

Date on which the reagent was manufactured or the bottle was opened.
For new titrants, the time the preparation was made will be specified.

Format: YYYY:MM:DD

Monitoring

on | off (Default value: off)

If this parameter is activated, then the working life will be monitored.

Working life

If you define a time interval for the working life, then the Expiry date will
be tracked automatically.

Input range

1 - 999 days

Default value

999 days

Expiry date

If you define an expiry date, then the Working life will be tracked auto-
matically.

Format: YYYY:MM:DD

Action

Selection of the action which is carried out when the time interval has
expired.

Selection

Display message | Document message | Cancel
determination

Default value

Display message
For all three options it is documented in the deter-
mination data (see dialog More determination
data / Messages
), that the time interval has been
expired.