28 Parameters

915 KF Ti-Touch

■■■■■■■■

267

off
No drift correction takes place.

Drift value

This parameter can only be edited with Drift correction = manual.

Drift for manual drift correction.

Input range

0.0 - 99.9 µL/min

Default value

0.0 µL/min

Measured value display

on | off (Default value: off)

If this parameter is activated, the currently measured value is displayed
during the conditioning.

Conditioning options [Cond. options]

Cond. stop volume

Maximum permissible volume that can be dosed during conditioning.
Conditioning is stopped when the specified volume is dosed. If condition-
ing is continued by pressing [

] once again, then the titrant volume that

has already been dosed will not be taken into account; i.e. the dosing
starts again at zero. The stop volume should be adjusted to the size of the
titration cell in order to prevent any overflow.

Input range

0.00000 - 9999.99 mL

Default value

20.0000 mL

Selection

off

Cond. stop time

Maximum permissible time over which conditioning may take place. Con-
ditioning is stopped when the specified time has elapsed.

Input range

1 - 999999 s

Selection

off

Default value

off

Delay reconditioning

on | off (Default value: off)

If this parameter is activated, the conditioning will not start before all the
method commands have been processed. Otherwise, the conditioning will
immediately start after the titration command.