30.7 Karl Fischer titration

■■■■■■■■■■■■■■■■■■■■■■

336

■■■■■■■■

915 KF Ti-Touch

Problem

Cause

Remedy

A side reaction is taking
place.

Use special reagents.

Adjust the method (increase/decrease the
temperature, external extraction).

See technical literature.

The pH value is not in the
optimum range anymore.

Add buffer, see technical literature.

The titration will not
be finished.

The titration cell is leaking.

Check seals and septum. If necessary,
replace.

Replace the molecular sieve.

The minimum increment is
too low.

Define Titration rate = user and increase the
minimum volume increment (Min. incre-
ment
) (see Chapter 28.1.2, page 261).

The stop criterion is unsuit-
able.

Adjust the control parameters (see Chapter
28.1.2, page 261)
:

Increase the stop drift.

Select a short delay time.

See also: The drift becomes
greater after each titration.

The sample is over-
titrated.

The increments at the end
of the titration are too
high.

Define Titration rate = user and reduce
the dosing rate (Max. rate) (see Chapter
28.1.2, page 261)
.
The following experiment provides a clue
for the optimum dosing rate: During condi-
tioning, display the drift and add sample
without starting the titration. Select a value
below the highest drift as dosing rate.

Stir faster.

The amount of methanol in
the working medium is too
low.

Replace the working medium.

Reduce the amount of solubility promoter,
if working with solvent mixtures, see tech-
nical literature.

The electrode may be cov-
ered.

Wipe off the electrode with ethanol or a suit-
able solvent.

The solution
becomes darker
after each titration.

Replace the working medium.

The electrode may be cov-
ered.

Wipe off the electrode with ethanol or a suit-
able solvent.