28.3 Evaluations (EVAL)

■■■■■■■■■■■■■■■■■■■■■■

282

■■■■■■■■

915 KF Ti-Touch

Upper limit

Measured value, time or volume for the upper limit.

The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the 32 Technical specifications chapter, page 355.

Measured value (measuring mode Ipol):
Input range

–2,000.0 - 2,000.0 mV

Default value

2,000.0 mV

Measured value (measuring mode Upol):
Input range

–200.00 - 200.00 µA

Default value

200.00 µA

Measured value (measuring mode T):
Input range

–20.0 - 150.0 °C

Default value

150.0 °C

Time:
Input range

0.0 - 999,999.9 s

Default value

999,999.9 s

Volume:
Input range

0.00000 - 9,999.99 mL

Default value

9,999.99 mL

28.3.4

Rate evaluation (EVAL RATE)

The rate evaluation provides the median dosing rate in a defined range
(so-called window). The median dosing rate is determined by means of lin-
ear regression through a minimum of three measuring points. If the dos-
ing is carried out with only one dosing device and if more than one cylin-
der volume is dosed, then the median dosing rate will be reduced as a
result of the filling time.

When the command is opened for the first time, a window over the entire
time range is already defined. In contrary to the potentiometric evaluation
of a DET/MET titration the individual windows may overlap. A maximum
of 9 windows can be defined.

[New]

Define a new window.

[Delete]

Delete the selected window.