30 troubleshooting, 1 editing methods, 2 sample table – Metrohm 915 KF Ti-Touch User Manual

Page 347: 3 results/statistics

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

30 Troubleshooting

915 KF Ti-Touch

■■■■■■■■

333

30 Troubleshooting

30.1

Editing methods

Problem

Cause

Remedy

The method com-
mand cannot be
inserted.

The method command is
disabled.

In the dialog Dialog options / Command
list
, activate the desired command (see "Block-
ing unneeded commands and fixed keys",
page 34)
.

No control device which
supports the method com-
mand is configured in the
device manager.

Add a control device which supports the
desired method command to the list of devices
(see Chapter 11, page 74).

The additional eval-
uations (EVAL com-
mands) cannot be
inserted.

A titration or measuring
command to which the
evaluation can be applied
is missing.

1. Insert a titration or measuring command.
2. Insert the EVAL command directly after it.

An EVAL command
is depicted in red in
the command list.

The associated titration or
measuring command has
been deleted or replaced
by a different one.

1. Delete the EVAL command.
2. Insert a titration or measuring command.
3. Insert the EVAL command directly after it.

30.2

Sample table

Problem

Cause

Remedy

The method cannot
be entered in the
sample table.

The option Use sample
assignment table
is acti-
vated. The method
assigned to the sample
identification will be loa-
ded automatically at the
start of the determination.

Deactivate the option (see Chapter 15.1, page
139)
.

30.3

Results/Statistics

Problem

Cause

Remedy

No statistics are car-
ried out for a result.

Statistics is not activated.

In the method options, activate the option
Statistics (see Chapter 16.5, page 159).

Advertising