16.5 Method options

■■■■■■■■■■■■■■■■■■■■■■

162

■■■■■■■■

915 KF Ti-Touch

Selection

no | Selection of configured sample racks

Default value

no

Increase sample variable automatically

on | off (Default value: on)

The sample variable describes the current position of the sample on the
rack of the Sample Processor. If this option is activated, then the sample
variable is automatically increased by 1 at the end of each determination.

This option must be deactivated if the sample variable is to be altered spe-
cifically with a SAMPLE command.

Request rack position at start

on | off (Default value: off)

If this option is activated, then the rack position of the first sample to be
processed is requested.

16.5.2

Stop options

Main dialog: Edit parameters

Method options Start/Stop

options

Stop options

In the dialog Method options / Stop options, you can define the
actions which are carried out when a method is canceled.

The method can be canceled as follows:

Manual stop with the fixed key [

]

Stop because of an error

Stop by remote signal via the Control Remote Box

Switch off pumps

on | off (Default value: on)

If this option is activated, then all of the connected pumps will be
switched off.

Switch off stirrers

on | off (Default value: on)

If this option is activated, then all of the connected stirrers will be
switched off.

Set remote lines

Selection of the signal out of the templates or entering the required bit
pattern. Templates are defined under System

Templates Output

lines.