28 Parameters

915 KF Ti-Touch

■■■■■■■■

319

28.8

Automation

28.8.1

Rotating sample rack (MOVE)

The command MOVE is used to move to a rack position or an external
position.

Tower

Selection of the tower, with which the command is carried out. Both
Tower 1 and Tower 2 can always be selected, even if your Sample Pro-
cessor has only one tower.

Selection

1 | 2

Default value

1

1
View from the front, the right tower.
2
View from the front, the left tower.

Destination

Selection of the required target position.

Selection

Sample | Ext. position | Special beaker | Rack
position
| Current sample + | Current sample - |
Next position | Prev. position | Calibration
pos.
| rotate + | rotate - | swing + | swing -

Default value

Sample

Sample
Rack position, which is defined by the sample variable. The sample vari-
able is automatically increased by 1 after every method run or it can
selectively be modified with the command SAMPLE.
Ext. position
One of the four external positions which can be moved to with the
Swing Head. Enter the number of the external position (1…4) in the
field to the right.
Special beaker
Special beakers, which are defined in the rack table, can be directly
moved to. Enter the number of the special beaker (1…16) in field to
the right.
Rack position
Any rack position. Enter the number of the rack position (1…999) in
the field to the right.