Metrohm 915 KF Ti-Touch User Manual

Page 334

Advertising
background image

28.8 Automation

■■■■■■■■■■■■■■■■■■■■■■

320

■■■■■■■■

915 KF Ti-Touch

Current sample +
Starting from the current sample (defined by the sample variable) the
rack is moved forward by the number of rack positions (1…999)
entered in the field to the right.
Current sample -
Starting from the current sample (defined by the sample variable) the
rack is moved backward by the number of rack positions (1…999)
entered in the field to the right.
Next position
Starting from the current rack position, the rack is moved forward by
one position.
Prev. position
Starting from the current rack position, the rack is moved backward by
one position.
Calibration pos.
For automatic calibrations with a USB Sample Processor (relevant only
if the Ti-Touch supports calibrations).
rotate +
Moving the rack forward by a certain increment. The rotation incre-
ment is defined in the properties of the tower.
rotate -
Moving the rack backward by a certain increment. The rotation incre-
ment is defined in the properties of the tower.
swing +
Swinging the robotic arm outward by a certain increment (towards
higher angular degrees). The swing increment is defined in the proper-
ties of the Swing Head.
swing -
Swinging the robotic arm towards the center of the rack by a certain
increment (towards lower angular degrees). The swing increment is
defined in the properties of the Swing Head.

Beaker test action

This parameter can only be edited with Destination = Sample, Next
position
or Prev. position.

Selection of the action which is carried out, when the beaker sensor does
not detect any vessel at the sample position moved to. Additionally, ena-
ble the beaker sensor in the rack table.

Selection

Rotate rack | Display message

Default value

Display message

Advertising