8.4 Dosing unit

■■■■■■■■■■■■■■■■■■■■■■

52

■■■■■■■■

915 KF Ti-Touch

Cylinder volume

Cylinder volume of the dosing unit. It will be read out automatically on
dosing units with integrated data chips.

Selection

2 | 5 | 10 | 20 | 50

Default value

20

Serial no. cyl.

Serial number of the dosing cylinder. It will be read out automatically on
exchange or dosing units with integrated data chips. The number can be
changed manually at any time, e.g. when a cylinder is replaced.

Entry

8 digits maximum

[Valve disk]

Specify the shift direction of the valve disc (see Chapter 8.4.3, page 56).

[GLP test]

Define the time interval for the GLP test (see Chapter 8.6, page 61).

[PREP param.]

Enter the parameters for the preparation (see Chapter 8.4.1, page 52).

[Tubing param.]

Enter the parameters for the connected tubings (see Chapter 8.4.2, page
53)
.

8.4.1

Parameters for preparing (PREP) and emptying (EMPTY)

Titrant: Edit

Dosing unit PREP param.

In the dialog Dosing unit / PREP parameters, you can adjust the
parameters for the execution of the Prepare (command PREP) and
Empty (command EMPTY) functions. The Prepare function is used to
rinse the cylinder and tubing of the dosing unit and fill it air bubble-free.
You should carry out this function before the first determination or once
per day. The EMPTY function empties the cylinder and the tubings of the
dosing unit.

Dosing port PREP/EMPTY

Dosing port through which the cylinder contents are ejected.

Selection

Dosing port 1 | Dosing port 2 | Fill port | Spe-
cial port

Default value

Dosing port 1

Dosing rate Dos. port 1

Rate used for the aspiration and ejection of the reagent via dosing port 1.

Input range

0.01 - 166.00 mL/min