29 operation and maintenance, 1 system initialization – Metrohm 915 KF Ti-Touch User Manual

Page 345

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

29 Operation and maintenance

915 KF Ti-Touch

■■■■■■■■

331

29 Operation and maintenance

29.1

System initialization

In very rare instances, a faulty file system (e.g. because of a program
crash) may lead to an impairment of program functioning. The internal file
system must be initialized in such cases.

CAUTION

If you carry out a system initialization, then all user data except the
methods will be deleted.

We recommend creating a backup of the system at regular intervals in
order to avoid data losses.

Proceed as follows for the system initialization:

1 Switching off the instrument

Press the power switch at the rear.

2 Switching on the instrument

Press the power switch at the rear.
An image with a frog will be displayed after a couple of seconds.

While the image is being displayed, press the power switch again
and hold it down.

Do not release the power switch until an acoustic signal sounds.

A system message is displayed.

3 Confirm system initialization

Confirm the message with [Yes].
A second safety prompt is displayed.

Confirm this message with [Yes] as well.

Initialization is started. After the initialization has been completed,
the 915 KF Ti-Touch is started automatically.