ProSoft Technology MVI56-AFC User Manual

Page 232

Advertising
background image

MVI56-AFC Sample Logic

MVI56-AFC ♦ ControlLogix Platform

User Manual

Liquid and Gas Flow Computer

Page 232 of 316

ProSoft Technology, Inc.

February 25, 2011

Refer to the Meters[x].Analysis controller tag in order to move the concentrations

for each meter (x assumes values between 0 and 15).

The concentrations are entered as scaled integer format where 10000 = 100%.
For example:

C1 = 9168

≥ 91.68%

The sum of all concentration should be 100%. The chromatograph can measure

values which total is slightly less (or more) than 100%. In this case you should

configure the Normalization Error Tolerance parameter in the AFC Manager in

order to make sure that the module will not generate any alarms.
When the module detects that a molar concentration value has changed it will

generate an event. However, when the values are updated from ladder logic

using a chromatograph device it is not convenient to generate an alarm every

time a concentration value changes. In this case you may configure the module

to not generate any alarms when a molar value is modified.

Advertising
This manual is related to the following products: