MVI56-AFC ♦ ControlLogix Platform

Meter Proving

Liquid and Gas Flow Computer

User Manual

ProSoft Technology, Inc.

Page 49 of 316

February 25, 2011

4.1.5 Prover Characteristics

Prover Characteristics will vary based on the type of prover and options you

select. The following topics describe each field and its operating range.

Prover size units (65018.L)
This parameter sets the units in which the prover's base volume is represented.

This parameter is not meaningful for master meter provers.

Meter factor precision (65028+)
This parameter is a number between 0.00000001 and 0.0001. The default setting

is 0.0001

Pulse interpolation ratio (65030+)
Meter-proving pulse counts delivered to the AFC may be fractional, such as

when double chronometry is used with a SVP. This value is the number of

delivered counts that constitute a single actual pulse, so that the actual pulse

count is determined by dividing the delivered count by this. The default value is

1000.0 for compact provers and 1.0 for other types. This parameter is meaningful

only for non-master meter provers.

Flow tube linear coefficient of thermal expansion (65032+)
Holds the coefficient of thermal expansion of the prover barrel material,

meaningful only for non-master-meter provers. Here are some typical materials

and their expansion coefficients.
 Stainless steel 304 or 316

9.3e-6/°F

16.7e-6/°C

 Monel

7.9e-6/°F

14.3e-6/°C

 Carbon steel

6.2e-6/°F

11.2e-6/°C

 Invar

.8e-6/°F

1.4e-6/°C

The default value is that of carbon steel, 6.2e-6/°F 11.2e-6/°C.

Switch bar linear coefficient of thermal expansion (65034+)
Holds the coefficient of thermal expansion of the external switch bar material,

meaningful only for non-master-meter provers with option External switch bar

(register 65011 bit 6) set. Here are some typical materials and their expansion

coefficients.
 Stainless steel 304 or 316

9.3e-6/°F

16.7e-6/°C

 Monel

7.9e-6/°F

14.3e-6/°C

 Carbon steel

6.2e-6/°F

11.2e-6/°C

 Invar

.8e-6/°F

1.4e-6/°C

The default value is that of invar

.8e-6/°F

1.4e-6/°C.