INFICON MDC-370 Thin Film Deposition Controller User Manual

Page 146

Advertising
background image

MDC-370 DEPOSITION CONTROLLER

Parameter Name

Length (bytes)

Allowable Range

Output #

1

(0-15)

Name

16

All ASCII Characters

Type

2

0 - 65535, least significant
byte first

Condition String

24

! | & ( )

I/O Card#

1

1-2

Output Pin#

1

2-9

Total 45 bytes

The Type parameter defines whether the output is undefined, user defined, or
internal. 0=Undefined, 65535=user defined. All other types are internal and are
created by the MDC for source pocket and sensor crystal position control.

When clearing an output you must set the type=0. When defining an output you
must set the type=65535 or FFh, FFh. Internal types can only be set by the MDC.

The condition string is described in the program I/O section of the manual. The
allowable characters are ! | & ( ). The individual conditions are represented by
numbers. The conditions are 1, 2, or 3 bytes each. The first byte is the condition
number. The second and third byte, if any, are the sub-condition number. The
allowable conditions are listed in the following table: