5 input, output and action setup – INFICON MDC-370 Thin Film Deposition Controller User Manual

Page 41

Advertising
background image

MDC-370 DEPOSITION CONTROLLER

b. BCD - With BCD format , the MDC-370 will create from one
to three outputs based on the number of crystals. For example, an
eight crystal sensor head will use three outputs. If crystal one is
the desired crystal, all outputs will be false. If crystal four is the
desired crystal, outputs one and two will be true and output three
will be false.

4.2.4.1 EXAMPLE USING MAXTEK’S RSH-600 SIX CRYSTAL SENSOR HEAD

The following is a list of the sensor parameter settings required to control
Maxtek’s RSH-600 six crystal sensor head.

Number of Crystals -

6

Shutter Relay Type

-

None

Control

-

Direct

Drive

-

Sngl Step

Feedback Type

-

Indiv

Rotator Delay

-

30

With the above parameter settings, the MDC-370 will create six position
feedback inputs called “SensorN CrystalX” where X ranges from 1 to 6
and one control output called SensorN Drive Up.

The inputs should be connected to the six position feedback pins on the
RSH-600 sensor head. Pin #1 of connector J1 on the sensor head should
be connected to the “SensorN Crystal1” input on the MDC-370. Pin #2 on
the sensor head should be connected to “SensorN Crystal2” on the MDC-
370 and so on. Pin #7 on the sensor head should be connected to pin #12
or any of the return pins when using the standard 370 I/O board. When
using the 370 Active I/O board then pin #7 of the sensor head should be
connected to one side of a 115VAC source. The other side of the
115VAC source should be connected to the other side of the six position
feedback inputs on the 370.

One pin from the SensorN Drive Up output should be connected to the
115 VAC voltage source and the other to J1. The remaining pin on J1
should connect to the other side of the 115 VAC power supply. In this
configuration, combined with the “single step” drive type, whenever the
MDC-360 needs to change crystals it will close the SensorN Drive Up
output for one second. This completes the circuit applying the 115 VAC
to the RSH-600.

4.2.5

INPUT, OUTPUT AND ACTION SETUP

The MDC-370’s inputs, outputs and actions can be used to provide control for, or
an interface to all sorts of vacuum system peripherals such as PLC system
controllers, substrate heaters, planetary rotators, etc. If your system doesn’t
require any special interfacing or control then you can skip to the next section.

PROGRAMMING AND CONTROLLER SETUP

4-9

Advertising