MDC-370 DEPOSITION CONTROLLER

26. Receive Output setup (Code #25)

Instructs the controller to enter the incoming Output parameters for Output #n into
memory. A description of the Output parameter list is given above.

Format: Header, Address, Instruction=25, Length=45,Output #(0-15), 2 byte
Type, 16 byte Name, 24 byte Condition string,1 byte Card#, 1 byte pin#,
Checksum

27. Send Action setup (Code #26)

Instructs the controller to send the Action parameter list for Action #n. A
description of the Action parameter list is as follows:

Parameter Name

Length (bytes)

Allowable Range

Action #

1

(0-15)

Action Type

1

(0-20)

Condition String

24

! | & ( )

Total 26 bytes

Format: Header, Instruction=26, Length=1, Action #(0-15), Checksum

The action type defines the action that will be taken when the condition is
evaluated as true. The following table contains the list of possible actions.

Action Name

Action Type

No Action

0

Manual Power

1

Zero Thickness

2

Reset Process

3

Abort Process

4

Halt Process

5

Terminate Deposit

6

Hold In State

7

Step From State

8

Sound Attention

9

Sound Alert

10

Sound Alarm

11

Start Process

12

Select Process 1-8

13-20

Switch Crystals

21

Start Timer #1-8

22-29

Start Timer #1-8

22-29

Set Softnode #1-8

30-37

Example: To instruct the controller to send the Action setup parameter list for
Action #14 the computer would send:

Chr$(255)+Chr$(254)+Chr$(1)+Chr$(26)+Chr$(1)+Chr$(14)+Chr$(214)

COMPUTER INTERFACE

11-18