INFICON MDC-370 Thin Film Deposition Controller User Manual

Page 152

Advertising
background image

MDC-370 DEPOSITION CONTROLLER

Value Description

Length

(bytes)

Format

Units

Power

4

String

%

Average Thickness

5

String

Average Rate

4

String

Å/sec

Active Crystal Health

2

String

%

Layer number

3

String

None

Rate deviation

4

String

%

Active process number

1

String (0-98)

Active material number

1

String (0-31)

Active source number

1

String (1-6)

Sensor #1-6 Rate

4x6

String

Å/sec

Sensor #1-6 Thickness

5x6

String

Sensor #1-6 Crystal Position

1x6

String (1-8)

Sensor #1-6 Crystal Health

2x6

String (0-99)

%

Source #1-6 Pocket Position

1x6

String (1-8)

Layer Time To Go

7

String

h:mm:ss

State Time To Go

7

String

h:mm:ss

Elapsed Process Time

7

String

h:mm:ss

Elapsed Layer Time

7

String

h:mm:ss

Elapsed State Time

7

String

h:mm:ss

Source #1-6 Status

1x6

Bit 0 = Shutter
Bit 1 = Rotator Drive Up
Bit 2 = Rotator Drive Down

Sensor #1-6 Status

1x6

Bit 0 = Shutter
Bit 1 = Rotator Drive Up
Bit 2 = Rotator Drive Down
Bit 3 = Enabled
Bit 5 = Failed

Controller State

1

Binary (0-25)

Abort Process Errors

1

Bit 6 = Max Rate&Min Pwr
Bit 7 = Min Rate&Max Pwr

Alarm 1 Errors

1

Bit 0 = Action Setup
Corrupted
Bit 1 =Output Setup
Corrupted
Bit 2 = Input Setup Corrupted
Bit 3 = Crystal Fail, Process
Halted
Bit 4 = Rate Establish Error
Bit 5 = Active Layer
Corrupted
Bit 6 = Active Process
Corrupted
Bit 7 = System Setup
Corrupted

COMPUTER INTERFACE

11-22

Advertising