Page 76
MDC-370 DEPOSITION CONTROLLER
Data Points/Minute
Approx. Storage Time
(minutes)
30
900
60
450
120
225
300
90
600
45
8. Time (00:00-23:59)
This parameter sets the system time. Time is entered in 24-hour format without a
digit separator “:”. For example, to enter 1:05 PM you must enter “1305”. This
parameter is only visible when the data logging option is installed.
9. Date (01/01/00-12/31/99)
This parameter sets the system date in month/day/year format. The complete date
must be entered without the digit separator “/” character, and with two digits for
each of the month, day and year. For example, to enter 5/2/94, you must enter
“050294”. This parameter is only visible when the data logging option is installed.
PROGRAMMING AND CONTROLLER SETUP