MDC-370 DEPOSITION CONTROLLER

Remote Action Code

Description

Action Trigger Type

0

Sound Alarm Action

Momentary

1

Sound Alert Action

Momentary

2

Sound Attention Action

Momentary

3

Step From State Action

One Shot

6

Clear Hold in State Action

Continuous

5

Set Hold in State Action

Continuous

6

Terminate Deposit Action

One Shot

7

Halt Process Action

One Shot

8

Switch Crystals

One Shot


All the momentary actions trigger the action for a 100-millisecond period for each
message received. The one shot actions trigger the action once for each message
received. For example, sending one step from state message would cause the
controller to step into the next state one time. The continuous actions trigger that
action indefinitely. So the set hold in state action will cause the controller to hold
in the current state until the clear hold in state action is received and the exit state
condition is meet.

Format: Header, Address=1, Instruction=38, Length=1, Action Code (0-
8), Checksum

Example: To initiate a terminate deposit action the computer would send:

Chr$(255)+Chr$(254)+Chr$(1)+Chr$(38)+Chr$(1)+Chr$(6)+Chr$(210)

40. Send run-time values (Code #39)

Instructs the controller to send one run-time value base on the value# received.

Format: Header, Address, Instruction=39, Length=1, value #(0-39) 1 byte, 1 byte
Checksum.

A description of the run-time value list is as follows:

COMPUTER INTERFACE 11-29