9 source parameters, 10 recorder parameters, 11 utility setup parameter – INFICON MDC-370 Thin Film Deposition Controller User Manual
Page 16: 12 other
MDC-370 DEPOSITION CONTROLLER
1.3.9
SOURCE PARAMETERS
Number of Pockets
1 to 8
Shutter Relay Type
Normally open, normally closed, or
none.
Shutter Delay
0.0 to 9.9 sec
Position Control
Manual, direct, BCD, or individual.
Position Drive
Up, down, Fast, inline, single step, or
double step.
Feedback Type
Individual, BCD, single home, in
position, or no feedback.
Rotator Delay
0 to 99 sec
Source Voltage Range
2.5, 5, 10 volts
1.3.10 RECORDER PARAMETERS
Recorder #1/#2 Output
Rate, rate dev., power or thickness
Recorder #1/#2 Scale
Full scale %, 2/3 digit
1.3.11 UTILITY SETUP PARAMETER
Crystal Frequency
2.5, 3, 5, 6, 9, 10 MHz
Simulate Mode
On/Off
Interface Address
1 to 32
Attention Volume
0 to 10
Alert Volume
0 to 10
Alarm Volume
0 to 10
Data Points/Minute
30,60,120,300,600 PPM
Time
0 to 23:59
Date
MM/DD/YY
1.3.12 OTHER
Input Power Requirements
100, 120, 200, 240 VAC; 50/60 Hz;
25 watts
Operating Temperature Range
0 to 50
°C
Physical Weight
10 LB
Physical Size
19” rackmount case
3 1/2” high x 9 3/8” deep
GENERAL DESCRIPTION