MDC-370 DEPOSITION CONTROLLER

and maintaining the power level, a large deposition thickness can be achieved
with one crystal. The primary sensor must have an individual shutter for the rate
sample feature. The default for this parameter is 100% which enables sampling at
all times.

35. Sample Period (0 to 9:59:59)

The Sample Period parameter defines the sample period. For example, a sample
time of 5 minutes and a dwell of 40% will result in the crystal being sampled for 2
minutes, then the crystal shutter is automatically closed for the remaining 3
minutes while the deposition power is kept constant. Please note, once the crystal
shutter has opened, there is a 5-second delay for crystal stabilization before
measuring. This parameter will not be displayed if this feature is disabled based
on the settings of the prior parameter.

36. Configure Sensors

Pressing the right arrow key from this point will bring you to the Configure
Sensors screen. In this screen you define which sensors are measured, the tooling
factors, etc. The following is a list the Configure Sensors parameters:

a. Tooling (10.0 to 499.9%)

This parameter sets the tooling factor for a particular sensor which is used to
compensate for geometric factors in the deposition system which result in a
difference between the deposition rate on the substrates and the rate on the
sensing crystal. This parameter is entered in percent units and 100% corresponds
to equal rates at the substrate and at the sensing crystal. To a first approximation
the tooling factor can be calculated using the following equation:

where:

dc= Distance from source to crystal.

ds= Distance from source to substrate.

Empirical calibration of the tooling factor is described in Section 10.6.2.

b. Weight (10.0 to 499.9)

This parameter sets the weight factor for a particular sensor. The Weight factor is
used to adjust the effect a sensor has on the average rate. Each sensor's affect on
the average rate is ratio of the sensor's weight divided by the total weight for all
enabled sensors. This parameter is irrelevant when only one sensor is enabled for
measurement. The default for this parameter is 100.0

c. Crystal Fail (NotUsed, Disable, HltLast, Halt, TimePwr, Switch)

Tooling

dc

ds

%

=



 ⋅

2

100

PROGRAMMING AND CONTROLLER SETUP