Flowserve Logix MD+ ValveSight User Manual

Page 211

Advertising
background image

ValveSight™ Diagnostics DTM Manual for Logix MD+ Positioner with HART®

FCD-

LGENSF0014-00

© Flowserve Corporation

211


Command < ADC Range

Definition: During Command Loop Calibration, the 0% signal was out

of the Analog to Digital Converter (ADC) range.

Possible Solutions: The system is designed to accept a signal that is

above 10 ADC counts. Recalibrate making sure the ADC counts are
above that limit.

Command > ADC Range

Definition: During Command Loop Calibration, the 100% signal was

out of the Analog to Digital Converter (ADC) range.

Possible Solutions: The system is designed to accept a signal that is

below 4085 ADC counts. Recalibrate making sure the ADC counts are
below that limit.

Analog Output Calibration Flags

AO Range Small

Definition: During an Analog Output Calibration the difference between

the milliamp signal at 0% and the milliamp signal at 100% was too
small.

Possible Solutions: Recalibrate making sure to use a larger difference

between signal limits. This notification can be cleared by briefly
pressing the QUICK-CAL/ACCEPT button.


Pressure Calibration Flags

Pressure Calibration Required

Definition: A Factory Pressure Calibration has not been performed.

Unlike a regular pressure sensor calibration, a Factory Pressure
Calibration saves the calibration values to memory, making them
available should a factory reset be performed. Proper pressure sensor
calibration is required for proper pressure sensing and diagnostics.
Calibration values from a regular pressure sensor calibration will be lost
when a factory reset is performed. Typically no pressure calibration is
required with a new positioner.

Possible Solutions: After replacing a main board or a pressure sensor

board, perform a Factory Pressure Calibration. To do this, see the
Pressure Sensor Board Removal and Installation section of the IOM.

Pressure Board Fail Warning

Definition: One or more pressure sensors may have failed.

Possible Solutions: Check the supply pressure to ensure it is between

1.3 and 10.3 bar (19 and 150 PSI). Check the pressure sensor board
connections. Recalibrate the pressure sensors. If the problem persists,

Advertising
This manual is related to the following products: