Management Annunciator

The Management Annunciator page displays all modes, setup, configuration and calibration
errors and their current status. If a particular indicator is not wanted then a check in the box
for a particular item will mask that item and it will not be shown in the active indicator field at
the top of each page nor will it affect the status of the health bar on the main dashboard
view.

Status Area
The Active Indicator area shows the status of the most relevant active indicator. The color
of the "LED light" corresponds to the Active Indicator and the first color of the blink code
sequence on the positioner. Generally green indicates no immediate issues. Yellow
indicates a developing issue. Red indicates the ability to control may be compromised. A
detailed list of the indicators is given in the

Alarm Congfiguraion - Alarm Annunicator

page.

The Command Source field indicates weather the positioner is being controlled by digital or
an analog (4-20 mA) command source.